Improving CBCT Quality to CT Level using Deep-Learning with Generative Adversarial Network.

Researchers

Journal

Modalities

Models

Abstract

To improve image quality and CT number accuracy of daily cone-beam computed tomography (CBCT) through a deep-learning methodology with Generative Adversarial Network.
150 paired pelvic CT and CBCT scans were used for model training and validation. An unsupervised deep-learning method, 2.5D pixel-to-pixel generative adversarial network (GAN) model with feature mapping was proposed. A total of 12000 slice pairs of CT and CBCT were used for model training, while 10-cross validation was applied to verify model robustness. Paired CT-CBCT scans from an additional 15 pelvic patients and 10 head-and-neck (HN) patients with CBCT images collected at a different machine were used for independent testing purpose. Besides the proposed method above, other network architectures were also tested as: 2D vs. 2.5D; GAN model with vs. without feature mapping; GAN model with vs. without additional perceptual loss; and previously reported models as U-net and cycleGAN with or without identity loss. Image quality of deep-learning generated synthetic CT (sCT) images were quantitatively compared against the reference CT (rCT) image using mean absolute error (MAE) of Hounsfield units (HU) and peak signal-to-noise ratio (PSNR). The dosimetric calculation accuracy was further evaluated with both photon and proton beams.
The deep-learning generated synthetic CTs (sCT) showed improved image quality with reduced artifact distortion and improved soft tissue contrast. The proposed algorithm of 2.5 Pix2pix GAN with feature matching (FM) was shown to be the best model among all tested methods producing the highest PSNR and the lowest MAE to reference CT (rCT) . The dose distribution demonstrated a high accuracy in the scope of photon based planning, yet more work is needed for proton based treatment. Once the model was trained, it took 11-12 ms to process one slice, and could generate a 3D-volume of dCBCT (80 slices) in less than a second using a NVIDIA GeForce GTX Titan X GPU (12GB, Maxwell architecture).
The proposed deep-learning algorithm is promising to improve CBCT image quality in an efficient way, thus has a potential to support online CBCT-based adaptive radiotherapy.
This article is protected by copyright. All rights reserved.