Radiation dose reduction using deep learning-based image reconstruction for a low-dose chest computed tomography protocol: a phantom study.

Researchers

Journal

Modalities

Models

Abstract

The aim of this study was to compare the dose reduction potential and image quality of deep learning-based image reconstruction (DLIR) with those of filtered back-projection (FBP) and iterative reconstruction (IR) and to determine the clinically usable dose of DLIR for low-dose chest computed tomography (LDCT) scans.Multi-slice computed tomography (CT) scans of a chest phantom were performed with various tube voltages and tube currents, and the images were reconstructed using seven methods to control the amount of noise reduction: FBP, three stages of IR, and three stages of DLIR. For subjective image analysis, four radiologists compared 48 image data sets with reference images and rated on a 5-point scale. For quantitative image analysis, the signal to noise ratio (SNR), contrast to noise ratio (CNR), nodule volume, and nodule diameter were measured.In the subjective analysis, DLIR-Low (0.46 mGy), DLIR-Medium (0.31 mGy), and DLIR-High (0.18 mGy) images showed similar quality to the FBP (2.47 mGy) image. Under the same dose conditions, the SNR and CNR were higher with DLIR-High than with FBP and all the IR methods (all P<0.05). The nodule volume and size with DLIR-High were significantly closer to the real volume than with FBP and all the IR methods (all P<0.001).DLIR can improve the image quality of LDCT compared to FBP and IR. In addition, the appropriate effective dose for LDCT would be 0.24 mGy with DLIR-High.2023 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Similar Posts

Leave a Reply

Your email address will not be published. Required fields are marked *