Improving Image Quality and Reducing Scan Time for Synthetic MRI of Breast by Using Deep Learning Reconstruction.

Researchers

Journal

Modalities

Models

Abstract

To investigate a deep learning reconstruction algorithm to reduce the time of synthetic MRI (SynMRI) scanning on the breast and improve the image quality.A total of 192 healthy female volunteers (mean age: 48.1 years) underwent the breast MR examination at 3.0 T from September 2020 to June 2021. Standard SynMRI and fast SynMRI scans were collected simultaneously on the same volunteer. Deep learning technology with a generative adversarial network (GAN) was used to generate high-quality fast SynMRI images by end-to-end training. Peak signal-to-noise ratio (PSNR), mean squared error (MSE), and structural similarity index measure (SSIM) were used to compare the image quality of generated images from fast SynMRI by deep learning algorithms.Fast SynMRI acquisition time is half of the standard SynMRI scan, and the generated images of the GAN model show that PSNR and SSIM are improved and MSE is reduced.The application of deep learning algorithms with GAN model in breast MAGiC MRI improves the image quality and reduces the scanning time.Copyright © 2022 Jian Li et al.

Similar Posts

Leave a Reply