Deep learning based correction of low performing pixel in computed tomography.

Researchers

Journal

Modalities

Models

Abstract

Low Performing Pixel (LPP)/bad pixel in CT detectors cause ring and streaks artifacts, structured non-uniformities and deterioration of the image quality. These artifacts make the image unusable for diagnostic purposes. A missing/defective detector pixel translates to a channel missing across all views in sinogram domain and its effect gets spill over entire image in reconstruction domain as artifacts. Most of the existing ring and streak removal algorithms perform correction only in the reconstructed image domain. In this work, we propose a supervised deep learning algorithm that operates in sinogram domain to remove distortions cause by the LPP. This method leverages CT scan geometry, including conjugate ray information to learn the interpolation in sinogram domain. While the experiments are designed to cover the entire detector space, we emphasize on LPPs near detector iso-center as these have most adverse impact on image quality specially if the LPPs fall on the high frequency region (bone-tissue interface). We demonstrated efficacy of the proposed method using data acquired on GE RevACT multi-slice CT system with flat-panel detector. Experimental results on head scans show significant reduction in ring artifacts regardless of LPP location in the detector geometry. We have simulated isolated LPPs accounting for 5% and 10% of total channels. Detailed statistical analysis illustrates approximately 5dB improvement in SNR in both sinogram and reconstruction domain as compared to classical bicubic and Lagrange interpolation methods. Also, with reduction in ring and streak artifacts, the perceptual image quality is improved across all the test images.© 2022 IOP Publishing Ltd.

Similar Posts

Leave a Reply

Your email address will not be published. Required fields are marked *